FDAFDA recall Z-0083-2027 · Oct 14, 2026
Beckman Coulter Inc.: DxC 500i Clinical Analyzer Modules: 1 recall
Recall found
Product
DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU Module, DxC 500i, Part No. C63521 3. Access 2 Module, DxC 500i, Part No. C13252
Reason for recall
Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and [email protected] in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- Software version 1.3.5 or earlier. 1. UDI-DI 14987666545089 2. UDI-DI 14987666545072 3. UDI-DI 15099590742331
- Quantity
- 146 US, 170 OUS
- Distribution
- Worldwide - US Nationwide distribution.
- Recalling firm
- Beckman Coulter Inc., Brea, CA
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-09-02
Source: openFDA enforcement record Z-0083-2027 · search recall number Z-0083-2027 in FDA's Enforcement Report.
Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.