FDAFDA recall Z-0084-2027 · Oct 14, 2026

Beckman Coulter Inc.: DxC 500 AU Clinical Chemistry Analyzers: 1 recall

Recall found

Product

DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520

Reason for recall

Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and [email protected] in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
Software version 1.4.5 or earlier. 1. UDI-DI 14987666545058 2. UDI-DI 14987666545065
Quantity
146 US, 170 OUS
Distribution
Worldwide - US Nationwide distribution.
Recalling firm
Beckman Coulter Inc., Brea, CA
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-09-02

Source: openFDA enforcement record Z-0084-2027 · search recall number Z-0084-2027 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.