FDAFDA recall Z-0143-2027 · Oct 21, 2026
Boston Scientific Corporation: NaviFlex LW Pusher recall
Recall found
Product
NaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;
Reason for recall
Device was incorrectly manufactured with a guidewire exit port (referred to in the IFU as the RX port), which is not a design feature associated with this device configuration.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- Material Number (UPN): M00535030; UDI-DI: 08714729856191; Lot Number: 38904309, 39017263, 39365322;
- Quantity
- 403
- Distribution
- US Nationwide distribution in the states of AL, CA, FL, GA, IL, IN, KY, MD, NC, NH, NM, NY, RI, TX, WA, WI, WV and the countries in TBD.
- Recalling firm
- Boston Scientific Corporation, Marlborough, MA
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-09-10