Skip to content

Recall watch for the stuff you own · As an Amazon Associate we earn from qualifying purchases · Some links earn a commissionDisclosure

ConsumerSun

FDAFDA recall Z-0160-2027 · Oct 21, 2026

Beckman Coulter, Inc.: DxU 840m Iris recall

Recall found

Product

DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513

Reason for recall

The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
Product/UDI-DI: DxU 840m/15099590748654, DxU 850m/15099590748647. All Serial Numbers
Quantity
2,088
Distribution
Worldwide - US Nationwide distribution including in the states and territory of CA, NE, FL, NY, MA, MD, SC, WI, TX, OH, VA, PA, IL, NH, KS, GA, NC, IA, PR, TN, OR, UT, NJ, MI, CO, NM, MS, LA, RI, OK, WA, MO, ME, IN, CT, AZ, AR, MT, MN, ID, DE, AL, WV, DC, KY, NV, WY, ND, HI, AK, SD, VT, GU and the countries of Canada, Mexico, Switzerland, Panama, Ecuador, Japan, Hungary, Italy, Belgium, China, Germany, France, United Kingdom of Great Britain and Northern Ireland, Russian Federation, Singapore, Romania, Argentina, Israel, Spain, Belarus, Guatemala, Colombia, Qatar, Czechia, Reunion, Australia, United Arab Emirates, Poland, Croatia, Brazil, Slovakia, New Zealand, Saudi Arabia, Thailand, Republic of Korea, Serbia, Austria, Macao, Ukraine, India, Netherlands, Taiwan, South Africa, Sint Maarten, Bangladesh, Hong Kong, Viet Nam, Bahrain, Uganda, Lebanon, Morocco, Guadeloupe, Greece, Curacao, Republic of Moldova, Malaysia, Ghana, Indonesia, Lithuania, Portugal, Kazakhstan, Georgia, Philippines, Bosnia and Herzegovina, Libya, Kuwait, Oman, Jordan, Uzbekistan.
Recalling firm
Beckman Coulter, Inc., Miami, FL
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-08-12

Source: openFDA enforcement record Z-0160-2027 · search recall number Z-0160-2027 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 26, 2026, 8:14 PM EDT. The agency notice is the final word.