FDAFDA recall Z-0161-2027 · Oct 21, 2026
Beckman Coulter, Inc.: IQ200 SPRINT recall
Recall found
Product
IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684; IQ200 ELITE, REF: 700-3370, 700-3375, 700-3375RFB, 700-3376, B48999, C10683
Reason for recall
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- Product/UDI-DI: IQ200 SPRINT/10837461001256, IQ200 SELECT/10837461001300, IQ200 ELITE/10837461001355. All Serial Numbers
- Distribution
- Worldwide - US Nationwide distribution including in the states and territory of CA, NE, FL, NY, MA, MD, SC, WI, TX, OH, VA, PA, IL, NH, KS, GA, NC, IA, PR, TN, OR, UT, NJ, MI, CO, NM, MS, LA, RI, OK, WA, MO, ME, IN, CT, AZ, AR, MT, MN, ID, DE, AL, WV, DC, KY, NV, WY, ND, HI, AK, SD, VT, GU and the countries of Canada, Mexico, Switzerland, Panama, Ecuador, Japan, Hungary, Italy, Belgium, China, Germany, France, United Kingdom of Great Britain and Northern Ireland, Russian Federation, Singapore, Romania, Argentina, Israel, Spain, Belarus, Guatemala, Colombia, Qatar, Czechia, Reunion, Australia, United Arab Emirates, Poland, Croatia, Brazil, Slovakia, New Zealand, Saudi Arabia, Thailand, Republic of Korea, Serbia, Austria, Macao, Ukraine, India, Netherlands, Taiwan, South Africa, Sint Maarten, Bangladesh, Hong Kong, Viet Nam, Bahrain, Uganda, Lebanon, Morocco, Guadeloupe, Greece, Curacao, Republic of Moldova, Malaysia, Ghana, Indonesia, Lithuania, Portugal, Kazakhstan, Georgia, Philippines, Bosnia and Herzegovina, Libya, Kuwait, Oman, Jordan, Uzbekistan.
- Recalling firm
- Beckman Coulter, Inc., Miami, FL
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-08-12
Source: openFDA enforcement record Z-0161-2027 · search recall number Z-0161-2027 in FDA's Enforcement Report.
Copied from the agency record without changes. Data as of Oct 26, 2026, 8:14 PM EDT. The agency notice is the final word.