FDAFDA recall Z-2504-2026 · Jun 24, 2026
Medline Industries, LP: MEDLINE Medical Procedure Kits labeled as: 1) D AND C PACK recall
Recall found
Product
MEDLINE Medical Procedure Kits labeled as: 1) D AND C PACK, Medline Kit Number/SKU DYNJ53973C; 2) PK CUST CV A&B CABRINI CO, Medline Kit Number/SKU DYNJ60583C.
Reason for recall
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21KME435; Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21JMC247; Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21JMB678; Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21IMH919; Medline Product Number/SKU DYNJ60583C: UDI/DI 10195327605582 (each), 40195327605583 (case), Lot Number 24CBO389; Medline Product Number/SKU DYNJ60583C: UDI/DI 10195327605582 (each), 40195327605583 (case), Lot Number 24ABR872.
- Quantity
- 168 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
- Recalling firm
- Medline Industries, LP, Northfield, IL
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-05-04
Source: openFDA enforcement record Z-2504-2026 · search recall number Z-2504-2026 in FDA's Enforcement Report.
Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.