FDAFDA recall Z-2549-2026 · Jul 8, 2026
Baxter Healthcare Corporation: Hillrom VOLARA P recall
Recall found
Product
Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device
Reason for recall
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class I: FDA's most serious class, where use of or exposure to the product may cause serious health problems or death.
- Status
- Ongoing
- Codes, lots and dates
- UDI: 10887761985015; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
- Quantity
- 0
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Recalling firm
- Baxter Healthcare Corporation, Deerfield, IL
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-05-21
Source: openFDA enforcement record Z-2549-2026 · search recall number Z-2549-2026 in FDA's Enforcement Report.
Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.