FDAFDA recall Z-2552-2026 · Jul 1, 2026

Stryker Neurovascular: INZONE DETACHMENT SYSTEM recall

Recall found

Product

INZONE DETACHMENT SYSTEM, REF: M00345100950

Reason for recall

Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
UDI-DI: 04546540697950. Lots: WMP133638, WMP133871, WMP133872, WMP133988, WMP133989, WMP134016, WMP134132, WMP134133, WMP134234, WMP134235, WMP134255, WMP134256, WMP134380, WMP134381, WMP134432, WMP134433, WMP134518, WMP134519, WMP134609, WMP134610, WMP134794, WMP134795, WMP134939, WMP134940, WMP135187, WMP135243, WMP135302, WMP135314, WMP135354, WMP135355, WMP135387, WMP135417, WMP135473, WMP135475, WMP135491, WMP135492, WMP135623, WMP135720, WMP135727, WMP135807, WMP135865, WMP135897, WMP136084, WMP136107, WMP136190, WMP136200, WMP136611, WMP136642, WMP136736, WMP136748, WMP136852, WMP136894, WMP136985
Quantity
44,937
Distribution
Worldwide - US Nationwide distribution including in the states of PA, AZ, FL, CA, IL, AK, NY, MI, NJ, WI, SD, KY, AR, TX, AL, TN, MO, MA, IN, NE, WV, MT, CT, NC, VA, NV, CO, DC, DE, OH, ME, GA, WA, SC, GU, MN, OK, NH, MD, ND, UT, HI, OR, LA, MS, KS, NM, ID, IA, VT and the countries of ARGENTINA, AUSTRIA, BELGIUM, BRASIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, KOREA, LATVIA, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH PACIFIC, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM.
Recalling firm
Stryker Neurovascular, Fremont, CA
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-04-06

Source: openFDA enforcement record Z-2552-2026 · search recall number Z-2552-2026 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.