FDAFDA recall Z-2552-2026 · Jul 1, 2026
Stryker Neurovascular: INZONE DETACHMENT SYSTEM recall
Recall found
Product
INZONE DETACHMENT SYSTEM, REF: M00345100950
Reason for recall
Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- UDI-DI: 04546540697950. Lots: WMP133638, WMP133871, WMP133872, WMP133988, WMP133989, WMP134016, WMP134132, WMP134133, WMP134234, WMP134235, WMP134255, WMP134256, WMP134380, WMP134381, WMP134432, WMP134433, WMP134518, WMP134519, WMP134609, WMP134610, WMP134794, WMP134795, WMP134939, WMP134940, WMP135187, WMP135243, WMP135302, WMP135314, WMP135354, WMP135355, WMP135387, WMP135417, WMP135473, WMP135475, WMP135491, WMP135492, WMP135623, WMP135720, WMP135727, WMP135807, WMP135865, WMP135897, WMP136084, WMP136107, WMP136190, WMP136200, WMP136611, WMP136642, WMP136736, WMP136748, WMP136852, WMP136894, WMP136985
- Quantity
- 44,937
- Distribution
- Worldwide - US Nationwide distribution including in the states of PA, AZ, FL, CA, IL, AK, NY, MI, NJ, WI, SD, KY, AR, TX, AL, TN, MO, MA, IN, NE, WV, MT, CT, NC, VA, NV, CO, DC, DE, OH, ME, GA, WA, SC, GU, MN, OK, NH, MD, ND, UT, HI, OR, LA, MS, KS, NM, ID, IA, VT and the countries of ARGENTINA, AUSTRIA, BELGIUM, BRASIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, KOREA, LATVIA, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH PACIFIC, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM.
- Recalling firm
- Stryker Neurovascular, Fremont, CA
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-04-06
Source: openFDA enforcement record Z-2552-2026 · search recall number Z-2552-2026 in FDA's Enforcement Report.