FDAFDA recall Z-2553-2026 · Jul 1, 2026

GE Medical Systems SCS: GE Healthcare AW Server 3 recall

Recall found

Product

GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System

Reason for recall

GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
REF# (01)00840682102384(10)AWS03D02E6D5, GTIN Number 00840682102384
Quantity
340 servers
Distribution
Worldwide distribution - United States Nationwide and the countries of Algeria, Australia, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Morocco, Nepal, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, United Arab Emirates, United Kingdom, Vietnam.
Recalling firm
GE Medical Systems SCS, Buc, N/A
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-05-08

Source: openFDA enforcement record Z-2553-2026 · search recall number Z-2553-2026 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.

Watch my items

Get an email when new recalls are posted for the kinds of things you own. Just your first name and email, nothing else.

Unsubscribe any time with one click. We never share your address.