FDAFDA recall Z-2605-2026 · Jul 8, 2026
Insulet Corporation: Omnipod (Eros) Pod recall
Recall found
Product
Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).
Reason for recall
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class I: FDA's most serious class, where use of or exposure to the product may cause serious health problems or death.
- Status
- Ongoing
- Codes, lots and dates
- 1. Model/REF Number: POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72514. 2. Model/REF Number: ZXR425 (10-Pack), ZXR420 (Single Pod). Pack UDI-DI: N/A. Single Pod UDI-DI: N/A. Lot Numbers: L72516, L72521. 3. Model/REF Number: ZXP425 (10 Pack), ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72515, L72509, L72513, L72511, L72510, L71480, L72514.
- Quantity
- 289,046 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.
- Recalling firm
- Insulet Corporation, Acton, MA
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-05-20
Source: openFDA enforcement record Z-2605-2026 · search recall number Z-2605-2026 in FDA's Enforcement Report.
Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.