FDAFDA recall Z-2624-2026 · Jul 8, 2026
Medtronic Heart Valves Division: Medtronic Harmony Delivery Catheter System recall
Recall found
Product
Medtronic Harmony Delivery Catheter System
Reason for recall
Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a manufacturing issue.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class I: FDA's most serious class, where use of or exposure to the product may cause serious health problems or death.
- Status
- Ongoing
- Codes, lots and dates
- GTIN 00763000341367, Lot Numbers: 0012508285, 0012517045, 0012526102, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663878, 0012663879, 0012690218, 0012699635, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824770, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965210, 0012965211, 0012976400, 0012986693, 0013063766, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624, 0013236643, 0013272796, 0013272798. GTIN 00763000544027, Lot Numbers: 0012508285, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663879, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965211, 0012976400, 0012986693, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624. GTIN 00763000520151, Lot Numbers: 0012575897, 0012642965, 0012728368, 0013272799. GTIN 00763000582951, Lot Numbers: 0012590535, 0012606411, 0012606413, 0012642964, 0012659418, 0012678282, 0012690219, 0012699637, 0012804151, 0012804580, 0012874527, 0012888682, 0012896885, 0012965221, 0012986699, 0013063764, 0013063765, 0013105037, 0013227504. GTIN 00763000918712, Lot Numbers: 0012635392, 0012676200, 0012690207, 0012874528, 0012874529, 0013015539, 0013105038.
- Quantity
- 1881 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Canada, Chile, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Mexico, Netherlands, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
- Recalling firm
- Medtronic Heart Valves Division, Santa Ana, CA
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-05-22
Source: openFDA enforcement record Z-2624-2026 · search recall number Z-2624-2026 in FDA's Enforcement Report.
Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.