Skip to content

Recall watch for the stuff you own · As an Amazon Associate we earn from qualifying purchases · Some links earn a commissionDisclosure

Consumer Sun

FDAFDA recall Z-2651-2026 · Aug 5, 2026

PHILIPS MEDICAL SYSTEMS: IQon Spectral CT IQon Spectral CT Upgrades recall

Recall found

Product

IQon Spectral CT IQon Spectral CT Upgrades

Reason for recall

The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
FCO72800812; FCO72800814 728331 728332, DI (01)00884838059542
Quantity
42
Distribution
US Nationwide Distribution including Puerto Rico
Recalling firm
PHILIPS MEDICAL SYSTEMS, Cambridge, MA
Voluntary or mandated
FDA Mandated
Recall started
2026-06-11

Source: openFDA enforcement record Z-2651-2026 · search recall number Z-2651-2026 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.