FDAFDA recall Z-2679-2026 · Jul 22, 2026

Oscor Inc.: Abiomed REF: 0052-3021 recall

Recall found

Product

Abiomed REF: 0052-3021, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F, 23F. heart pump and accessories.

Reason for recall

Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class I: FDA's most serious class, where use of or exposure to the product may cause serious health problems or death.
Status
Ongoing
Codes, lots and dates
All Lots within Expiry/GTIN:00885672009786/00813502010558
Quantity
1332 units
Distribution
Distribution in the US state of MA
Recalling firm
Oscor Inc., Palm Harbor, FL
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-06-10

Source: openFDA enforcement record Z-2679-2026 · search recall number Z-2679-2026 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.

Watch my items

Get an email when new recalls are posted for the kinds of things you own. Just your first name and email, nothing else.

Unsubscribe any time with one click. We never share your address.