FDAFDA recall Z-2698-2026 · Jul 15, 2026
Medtronic MiniMed, Inc.: MiniMed 780G Insulin Pump recall
Recall found
Product
MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
Reason for recall
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523. Software version 6.62
- Quantity
- 778
- Distribution
- International distribution to the countries of Spain, Great Britain, Netherlands, Italy.
- Recalling firm
- Medtronic MiniMed, Inc., Northridge, CA
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-05-27
Source: openFDA enforcement record Z-2698-2026 · search recall number Z-2698-2026 in FDA's Enforcement Report.