FDAFDA recall Z-2715-2026 · Jul 15, 2026

Biomerieux Inc: VITEK 2 AST-N447 TEST KIT Product reference 424620 recall

Recall found

Product

VITEK 2 AST-N447 TEST KIT Product reference 424620

Reason for recall

Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
Lot 0853293604 UDI-DI 03573026634551 Product is not registered in the US. OUS only.
Quantity
45,900 cards
Distribution
International distribution to the countries of Belgium and the Netherlands.
Recalling firm
Biomerieux Inc, Hazelwood, MO
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-06-04

Source: openFDA enforcement record Z-2715-2026 · search recall number Z-2715-2026 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.

Watch my items

Get an email when new recalls are posted for the kinds of things you own. Just your first name and email, nothing else.

Unsubscribe any time with one click. We never share your address.