Skip to content

Recall watch for the stuff you own · As an Amazon Associate we earn from qualifying purchases · Some links earn a commissionDisclosure

Consumer Sun

FDAFDA recall Z-2727-2026 · Jul 22, 2026

GE Medical Systems, LLC: Optima XR220amx recall

Recall found

Product

Optima XR220amx, Mobile Digital-Ready Radiographic System

Reason for recall

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
UDI-DI: 00840682115261; System IDs: 05249MOB02.OLD, 205820DP1, 228818XR220, 229276M1, 240566P1, 301295OPTIMA2, 303788SBOPT1, 314251P5, 352LEEDP1, 386FLGDPXR3, 417347DPORT1, 512SAMCSB, 516632OPTIMA2, 540981ODP7, 563264OPT, 601445DP1, 615338HDP01, 661949DP2, 713441OPCDX1, 717657OPT1, 718206OPT240A, 727848SB1, 732321DP1, 757736DP3, 813302OPT1, 828213DXP1, 828286OPT1, 843847220D, 850070941, 865545DP3, 915821OPT1, 941342OP221, 972566XR220E, 985447TM4, SA1208MM05, SB2201700085WK, XR1034798WK7, XR1037005WK4
Quantity
38 units
Distribution
Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.
Recalling firm
GE Medical Systems, LLC, Waukesha, WI
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-06-05

Source: openFDA enforcement record Z-2727-2026 · search recall number Z-2727-2026 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.