FDAFDA recall Z-2771-2026 · Jul 29, 2026
Boston Scientific Corporation: Boston Scientific TheraSphere 360 Y-90 Management Platform recall
Recall found
Product
Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.
Reason for recall
Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- UDI/DI 00191506068264, All released software versions (V1.1.2, V1.2.1, V1.2.2. V2.0.1, V2.0.2)
- Quantity
- 68 units
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY.
- Recalling firm
- Boston Scientific Corporation, Maple Grove, MN
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-06-19
Source: openFDA enforcement record Z-2771-2026 · search recall number Z-2771-2026 in FDA's Enforcement Report.