FDAFDA recall Z-2811-2026 · Aug 5, 2026
Medtronic Perfusion Systems: EOPA 3D Arterial Cannula 22FR recall
Recall found
Product
EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78322
Reason for recall
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- UDI-DI: 20199150023251, Lot Numbers: 0232709184 0232745036 0232798416 0232862784 0232862787; UDI-DI: 00199150023257, Lot Numbers: 0232862787
- Quantity
- 1,424 units
- Distribution
- Worldwide - US Nationwide distribution.
- Recalling firm
- Medtronic Perfusion Systems, Brooklyn Park, MN
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-06-24
Source: openFDA enforcement record Z-2811-2026 · search recall number Z-2811-2026 in FDA's Enforcement Report.