FDAFDA recall Z-2859-2026 · Aug 12, 2026
Innovasis, Inc: Anti-Torque recall
Recall found
Product
Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System
Reason for recall
Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- LS-749: UDI-DI: M711BTA2020, M711LS7490, Serial Numbers: 002, 003, 007, 012, 014
- Quantity
- 5
- Distribution
- US Nationwide distribution in the states of TX, CA.
- Recalling firm
- Innovasis, Inc, Salt Lake City, UT
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-05-22
Source: openFDA enforcement record Z-2859-2026 · search recall number Z-2859-2026 in FDA's Enforcement Report.