FDAFDA recall Z-2958-2026 · Aug 26, 2026
Arrowhead DE LLC: Arrow-LOK Implant recall
Recall found
Product
Arrow-LOK Implant, REF: 21-253519, 21-253522, 21-253525, 21-253528, 21-253530, 21-253533, 21-253535, 21-253540, 21-253545, 21-253550, 21-353513, 21-353516, 27-253519, 27-253522, 27-253525, 27-253528, 27-253530, 27-253533, 27-253535. Implant indicated for fixation of osteotomies, arthrodeses and reconstruction in the lesser toes following corrective procedures.
Reason for recall
Implant did not undergo required gamma radiation dose audit. A potential reduction in sterility assurance could lead to patient infection.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- REF/UDI-DI/Lot: 21-253519/00855967006035/24662119, 242162; 21-253522/00855967006042/24662122, 242163, 220702; 21-253525/00855967006059/24662125, 220703, 242164; 21-253528/00855967006066/24662128, 242165, 230704; 21-253530/00855967006073/242166, 230705, 220705; 21-253533/00855967006080/24662133, 242167, 230706; 21-253535/00855967006097/24662135, 242168, 230707; 21-253540/00855967006110/242169; 21-253545/00855967006127/242170; 21-253550/00855967006134/24662150; 21-353513/00855967006004/24662113, 242160; 21-353516/00855967006011/24662116, 230701, 242161; 27-253519/00855967006028/24662719, 242171; 27-253522/00855967006141/242172; 27-253525/00855967006158/242173; 27-253528/00855967006165/242174; 27-253530/00855967006172/242175; 27-253533/00855967006189/24662733, 242176; 27-253535/00855967006196/24662735, 242177
- Quantity
- 4,171
- Distribution
- US Nationwide distribution in the states of VT, FL, TN, NY, GA, TX, HI.
- Recalling firm
- Arrowhead DE LLC, Peachtree City, GA
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-06-23
Source: openFDA enforcement record Z-2958-2026 · search recall number Z-2958-2026 in FDA's Enforcement Report.