Skip to content

Recall watch for the stuff you own · As an Amazon Associate we earn from qualifying purchases · Some links earn a commissionDisclosure

Consumer Sun

FDAFDA recall Z-2974-2026 · Aug 26, 2026

Beckman Coulter Inc.: Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1 recall

Recall found

Product

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers

Reason for recall

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065
Distribution
Worldwide distribution - US Nationwide.
Recalling firm
Beckman Coulter Inc., Brea, CA
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-07-15

Source: openFDA enforcement record Z-2974-2026 · search recall number Z-2974-2026 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.