Skip to content

Recall watch for the stuff you own · As an Amazon Associate we earn from qualifying purchases · Some links earn a commissionDisclosure

Consumer Sun

FDAFDA recall Z-2987-2026 · Aug 26, 2026

restor3d Inc.: Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid recall

Recall found

Product

Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder

Reason for recall

Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
Lot Code: Model No 1500-1030; UDI-DI 00840097513225; Lot Numbers A00030, A00065, A09828, A09830 Model No 1500-1035; UDI-DI 00840097513232; Lot Numbers A00064, A09829
Quantity
59
Distribution
U.S. Nationwide distribution in the state of Texas.
Recalling firm
restor3d Inc., Durham, NC
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-06-30

Source: openFDA enforcement record Z-2987-2026 · search recall number Z-2987-2026 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.