FDAFDA recall Z-2992-2026 · Sep 2, 2026
Medline Industries, LP: Medline breathing circuits labeled as: 1) HUDSON RCI recall
Recall found
Product
Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEONATAL,21 V, Medline Product Number/SKU HUD78009 (Teleflex Product Number/SKU 780-09); 3) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-13 (Teleflex Product Number/SKU 780-13); 4) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-14 (Teleflex Product Number/SKU 780-14); 5) HUDSON RCI, DBD-CIRCUIT, HTD, DUAL LIMB, NEONATE, Medline Product Number/SKU 780-15 (Teleflex Product Number/SKU 780-15); 6) HUDSON RCI, CIRCUIT,NEO,DUAL,WIRE,60 IN, Medline Product Number/SKU HUD78018) (Teleflex Product Number/SKU 780-18).
Reason for recall
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- 1) Medline Product Number/SKU HUD78007, UDI/DI 10197344042463 (each) 20197344042460 (case), All lot numbers; 2) Medline Product Number/SKU HUD78009, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 3) Medline Product Number/SKU 780-13, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 4) Medline Product Number/SKU 780-14, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 5) Medline Product Number/SKU 780-15, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 6) Medline Product Number/SKU HUD78018, UDI/DI 10197344010660 (each) 20197344010667 (case), All lot numbers.
- Quantity
- 145525 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Panama.
- Recalling firm
- Medline Industries, LP, Northfield, IL
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-07-13
Source: openFDA enforcement record Z-2992-2026 · search recall number Z-2992-2026 in FDA's Enforcement Report.
Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.
