FDAFDA recall Z-3087-2026 · Sep 9, 2026

Fresenius Medical Care Renal Therapies Group, LLC: 5008X HD/HDF with Twister Blood Tubing Set Part Number 03-5350-8 recall

Recall found

Product

5008X HD/HDF with Twister Blood Tubing Set Part Number 03-5350-8

Reason for recall

Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
Lots 25JR01802 25NR01017 25NR01266 UDI-DI 00840861102525
Quantity
6,624 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC, Waltham, MA
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-08-10

Source: openFDA enforcement record Z-3087-2026 · search recall number Z-3087-2026 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.

Watch my items

Get an email when new recalls are posted for the kinds of things you own. Just your first name and email, nothing else.

Unsubscribe any time with one click. We never share your address.