Skip to content

Recall watch for the stuff you own · As an Amazon Associate we earn from qualifying purchases · Some links earn a commissionDisclosure

Consumer Sun

FDAFDA recall Z-3130-2026 · Sep 23, 2026

Nova Biomedical Corporation: StatStrip Glucose and B- Ketone meter recall

Recall found

Product

StatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system.

Reason for recall

A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class I: FDA's most serious class, where use of or exposure to the product may cause serious health problems or death.
Status
Ongoing
Codes, lots and dates
Model No: 63683. UDI-DI: 00385480636834. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
Quantity
1,858 units
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, United Kingdom.
Recalling firm
Nova Biomedical Corporation, Waltham, MA
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-07-29

Source: openFDA enforcement record Z-3130-2026 · search recall number Z-3130-2026 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.

Watch my items

Get an email when new recalls are posted for the kinds of things you own. Just your first name and email, nothing else.

Unsubscribe any time with one click. We never share your address.