FDAFDA recall Z-3204-2026 · Sep 23, 2026
Howmedica Osteonics Corp.: TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from recall
Recall found
Product
TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Reason for recall
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- 620-00-28D UDI-DI 07613327313581 Lots MHANP2 MHD1N0
- Quantity
- 12,156 US
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
- Recalling firm
- Howmedica Osteonics Corp., Mahwah, NJ
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-07-30
Source: openFDA enforcement record Z-3204-2026 · search recall number Z-3204-2026 in FDA's Enforcement Report.