Skip to content

Recall watch for the stuff you own · As an Amazon Associate we earn from qualifying purchases · Some links earn a commissionDisclosure

Consumer Sun

FDAFDA recall Z-3309-2026 · Oct 7, 2026

Philips North America: Ingenia 3 recall

Recall found

Product

Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions: R11, R12 and R13;

Reason for recall

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
1. Model number (REF): 1. 781342 (3.0T); UDI-DI: 00884838055339; Serial Numbers: 71001, 71002, 71008, 71031, 71038, 71051, 71061, 71072, 71084, 71088, 71089, 71095, 71101, 71154, 71163, 71164, 71166, 71184, 71185, 71193, 71210, 71231, 71269, 71271, 71288, 71290, 71297, 71301, 71303, 71309, 71311, 71312, 71322, 71335, 71336, 71340, 71341, 71344, 71355, 71375, 71385, 71393, 71412, 71413, 71420, 71422, 71427, 71438, 71441, 71443, 71445, 71446, 71449, 71450, 71459, 71462, 71463, 71475, 71486, 71487, 71489, 71491, 71513, 71521, 71528, 71556, 71558, 71562, 71589, 71604, 71607, 71612, 71615, 71619, 71620, 71639, 71641, 71646, 71665, 71673, 71688, 71697, 71704, 71710, 71715, 71729, 71730, 71739, 71754, 71756, 71757, 71759, 71813, 71817, 71828, 71831, 71841, 71849, 71850, 71853, 71862, 71885, 71896, 71921, 71923, 71954, 71959, 71960, 71962, 71963, 71964, 71986, 71992, 85006, 85020, 85032, 85038, 85046, 85066, 85068, 85076, 85078, 85085, 85087, 85091, 85092, 85102, 85126, 85148, 85150, 85154, 85191, 85201, 85216, 85245, 85279, 85377; 2. Model Number (REF): 781377 (3.0T); UDI-DI: 00884838009813; Serial Numbers: 42012, 42014, 42015, 42017, 42033, 42036, 42048, 42059, 42088, 42092, 42109, 42125, 42126, 42127, 42146, 42169, 42196, 42198, 42206, 42210, 42218, 42220, 42224, 42228, 42234, 42238, 42247, 42258, 42265, 42271, 42282, 42288, 42293, 42305, 42307, 42321, 42322, 42330, 42342, 42352, 42353, 42359, 42360, 42386, 42389, 42396, 42403, 42409, 42412, 42416, 42425; 3. Model Number (REF): 782103 (3.0T); UDI-DI: 00884838098299; Serial Numbers: 85646, 85652, 85679, 85686, 85708, 85723, 85758, 85759, 85762; 4. Model Number (REF): 781271 (3.0T CX); UDI-DI: 00884838068452; Serial Numbers: 78008, 78009, 78011, 78012, 78017, 78018, 78027, 78028, 78033, 78035, 78036, 78039, 78040, 78042, 78045, 78053, 78060, 78062, 78063, 78067, 78068, 78070, 78095, 78100, 78103, 78112, 78113, 78115, 78122, 78145, 78147, 78148, 78155, 78166, 78167, 78186, 78195, 78201, 78202, 78208, 78215, 78217, 78229, 78277, 78286, 78287;
Quantity
243
Distribution
Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.
Recalling firm
Philips North America, Cambridge, MA
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-08-28

Source: openFDA enforcement record Z-3309-2026 · search recall number Z-3309-2026 in FDA's Enforcement Report.