FDAFDA recall Z-3324-2026 · Oct 7, 2026

Percussionaire Corporation: Travel Air recall

Recall found

Product

Travel Air, Product Number: S00055; Phasitron 5, Product Number: P5 used with Instruction Manual

Reason for recall

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
Product Number/UDI-DI: S00055/00849436000242, P5/00849436000501, Instruction Manual: P20106 Rev B
Quantity
228,596
Distribution
Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.
Recalling firm
Percussionaire Corporation, Sandpoint, ID
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-08-18

Source: openFDA enforcement record Z-3324-2026 · search recall number Z-3324-2026 in FDA's Enforcement Report.