SPECTRUM MEDICAL SRL: Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High F recall

Recall found

Product

Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Model/Catalog Number: HX55W-S0 Software Version: N/A Product Description: The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 8 l/min. Quantum PureFlow Standard Heat Exchanger is a single use device which allows heating/cooling the blood during extracorporeal circulation. The device is sold sterile in a blister pack in a multiple box of 4 units. Device surfaces in contact with blood are treated with a phosphorylcholine-based biocompatible coating. The device has been designed to be powered by heaters coolers that use water or a glycol-based heat transfer fluid (HTF). Component: N/A

Reason for recall

Due to blood-to-glycol leak in the heat exchanger.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
Lot Code: 18051160301178 (HX55W-S0)
Quantity
3232
Distribution
US Nationwide distribution in the states of AL, CA, FL, KS MA, MD, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WI.
Recalling firm
SPECTRUM MEDICAL SRL, Mirandola
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-07-17

Source: openFDA enforcement record Z-3328-2026 · search recall number Z-3328-2026 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.

Watch my items

Get an email when new recalls are posted for the kinds of things you own. Just your first name and email, nothing else.

Unsubscribe any time with one click. We never share your address.