FDAFDA recall Z-3012-2026 · Sep 2, 2026
Midmark Corporation: Midmark IQecg Digital ECG recall
Recall found
Product
Midmark IQecg Digital ECG, Model/Catalog Number: 4-000-0061, 4-000-0062 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.
Reason for recall
A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- Lot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark IQecg Digital ECG. The Midmark IQecg Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 480000 4820704 UDI-DI: 00841709100581, 00841709100321, 00841709100895, 00841709100901
- Quantity
- 3624 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Recalling firm
- Midmark Corporation, Versailles, OH
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-07-20
Source: openFDA enforcement record Z-3012-2026 · search recall number Z-3012-2026 in FDA's Enforcement Report.
Copied from the agency record without changes. Data as of Oct 22, 2026, 4:56 AM EDT. The agency notice is the final word.