FDAFDA recall D-0167-2027 · Nov 11, 2026

Fresenius Medical Care Renal Therapies Group, LLC: DELFLEX Peritoneal Dialysis Solution with 1 recall

Quick answer

Product
DELFLEX Peritoneal Dialysis Solution with 1.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay safe Exchange Set, 2,000 mL - Single Dose Bag, Sterile and Non-Pyrogenic, Rx only, CAT.NO. 054-20221. Fresenius Medical Care, Renal Therapies Group, LLC, Lawrence, MA 01843. NDC 46163-206-92
Brand
Fresenius Medical Care Renal Therapies Group, LLC
Codes, lots and dates
Lot #: 25JGPD045, Lot 25JGPD046, Lot 25JGPD060, Exp 1/31/2027; 25KGPD057, Exp 2/28/2027.
Hazard
Lack of Assurance of Sterility:complaints have been received related to potential leaks from small perforations in the print area on the primary bag in the affected lots.
Distribution
Nationwide in the USA.
Quantity
4,232 cases (5 bags per case)
Recall date
Nov 11, 2026
Agency
U.S. Food and Drug Administration (FDA) enforcement report
Recall number
D-0167-2027

Recall found

Product

DELFLEX Peritoneal Dialysis Solution with 1.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay safe Exchange Set, 2,000 mL - Single Dose Bag, Sterile and Non-Pyrogenic, Rx only, CAT.NO. 054-20221. Fresenius Medical Care, Renal Therapies Group, LLC, Lawrence, MA 01843. NDC 46163-206-92

Reason for recall

Lack of Assurance of Sterility:complaints have been received related to potential leaks from small perforations in the print area on the primary bag in the affected lots.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall facts
Quantity4,232 cases (5 bags per case)
Codes, lots and datesLot #: 25JGPD045, Lot 25JGPD046, Lot 25JGPD060, Exp 1/31/2027; 25KGPD057, Exp 2/28/2027.
DistributionNationwide in the USA.
Recall dateNov 11, 2026
ClassClass II
Agency IDD-0167-2027
Official noticeU.S. Food and Drug Administration (FDA) enforcement report
Data as of, last checkedNov 21, 2026, 3:48 PM EST

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
Lot #: 25JGPD045, Lot 25JGPD046, Lot 25JGPD060, Exp 1/31/2027; 25KGPD057, Exp 2/28/2027.
Quantity
4,232 cases (5 bags per case)
Distribution
Nationwide in the USA.
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC, Lawrence, MA
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-10-14

Source: openFDA enforcement record D-0167-2027 · search recall number D-0167-2027 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Nov 21, 2026, 3:48 PM EST. The agency notice is the final word.

FAQ

What was recalled?

Product: DELFLEX Peritoneal Dialysis Solution with 1.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay safe Exchange Set, 2,000 mL - Single Dose Bag, Sterile and Non-Pyrogenic, Rx only, CAT.NO. 054-20221. Fresenius Medical Care, Renal Therapies Group, LLC, Lawrence, MA 01843. NDC 46163-206-92 Brand: Fresenius Medical Care Renal Therapies Group, LLC Codes, lots and dates: Lot #: 25JGPD045, Lot 25JGPD046, Lot 25JGPD060, Exp 1/31/2027; 25KGPD057, Exp 2/28/2027.

What is the hazard?

Lack of Assurance of Sterility:complaints have been received related to potential leaks from small perforations in the print area on the primary bag in the affected lots.

How do I check if mine is affected?

Codes, lots and dates: Lot #: 25JGPD045, Lot 25JGPD046, Lot 25JGPD060, Exp 1/31/2027; 25KGPD057, Exp 2/28/2027.

Where was it distributed?

Nationwide in the USA.

Related recalls

Watch my items

Get an email when new recalls are posted for the kinds of things you own. Just your first name and email, nothing else.

Unsubscribe any time with one click. We never share your address.