FDAFDA recall Z-3225-2026 · Sep 30, 2026
Fresenius Medical Care Renal Therapies Group, LLC: CombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors recall
Quick answer
- Product
- CombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors; Model Number: 03-2522-1;
- Brand
- Fresenius Medical Care Renal Therapies Group, LLC
- Codes, lots and dates
- Model Number: 03-2522-1; UDI-DI: 00840861102402 (Bag), 10840861102409 (Case); Lot codes: 26AR01010, 26BR01072, 26CR01067, 26DR01160, 26DR01198, 26AR01011, 26BR01074, 26CR01069, 26DR01161, 26DR01203, 26AR01034, 26BR01076, 26CR01070, 26DR01162, 26DR01206, 26AR01035, 26BR01077, 26CR01072, 26DR01163, 26DR01207, 26AR01046, 26BR01078, 26CR01073, 26DR01165, 26DR01209, 26AR01047, 26BR01079, 26CR01074, 26DR01166, 26DR01210, 26AR01056, 26BR01082, 26CR01075, 26DR01167, 26DR01255, 26AR01057, 26BR01083, 26CR01076, 26DR01168, 26ER01022, 26AR01058, 26BR01084, 26CR01079, 26DR01170, 26ER01023, 26AR01059, 26BR01085, 26CR01080, 26DR01171, 26ER01024, 26AR01230, 26BR01087, 26CR01081, 26DR01172, 26ER01029, 26BR01007, 26BR01088, 26CR01082, 26DR01173, 26ER01034, 26BR01035, 26BR01210, 26CR01083, 26DR01174, 26ER01035, 26BR01036, 26CR01014, 26CR01084, 26DR01175, 26ER01037, 26BR01037, 26CR01045, 26CR01085, 26DR01177, 26ER01039, 26BR01039, 26CR01046, 26CR01086, 26DR01180, 26ER01040, 26BR01045, 26CR01047, 26CR01087, 26DR01182, 26ER01042, 26BR01049, 26CR01048, 26CR01088, 26DR01183, 26ER01044, 26BR01051, 26CR01049, 26CR01090, 26DR01184, 26ER01048, 26BR01052, 26CR01051, 26CR01091, 26DR01185, 26ER01049, 26BR01053, 26CR01052, 26CR01092, 26DR01186, 26ER01054, 26BR01057, 26CR01054, 26CR01093, 26DR01187, 26ER01055, 26BR01058, 26CR01056, 26CR01094, 26DR01188, 26ER01056, 26BR01059, 26CR01057, 26CR01206, 26DR01189, 26ER01062, 26BR01060, 26CR01058, 26CR01207, 26DR01190, 26ER01065, 26BR01063, 26CR01059, 26CR01208, 26DR01191, 26ER01066, 26BR01064, 26CR01060, 26CR01209, 26DR01193, 26ER01074, 26BR01065, 26CR01061, 26CR01210, 26DR01194, 26ER01266, 26BR01066, 26CR01064, 26DR01013, 26DR01195, 26ER01268, 26BR01067, 26CR01065, 26DR01158, 26DR01196, 26ER01282, 26BR01069, 26CR01066, 26DR01159, 26DR01197;
- Hazard
- Communication to recommend reviewing the IFU and ensuring that all treatment preparation and handling steps are performed as instructed.
- Distribution
- Domestic: Nationwide Distribution;
- Quantity
- 1743026 units
- Recall date
- Sep 30, 2026
- Agency
- U.S. Food and Drug Administration (FDA) enforcement report
- Recall number
- Z-3225-2026
Recall found
Product
CombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors; Model Number: 03-2522-1;
Reason for recall
Communication to recommend reviewing the IFU and ensuring that all treatment preparation and handling steps are performed as instructed.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
| Quantity | 1743026 units |
|---|---|
| Codes, lots and dates | Model Number: 03-2522-1; UDI-DI: 00840861102402 (Bag), 10840861102409 (Case); Lot codes: 26AR01010, 26BR01072, 26CR01067, 26DR01160, 26DR01198, 26AR01011, 26BR01074, 26CR01069, 26DR01161, 26DR01203, 26AR01034, 26BR01076, 26CR01070, 26DR01162, 26DR01206, 26AR01035, 26BR01077, 26CR01072, 26DR01163, 26DR01207, 26AR01046, 26BR01078, 26CR01073, 26DR01165, 26DR01209, 26AR01047, 26BR01079, 26CR01074, 26DR01166, 26DR01210, 26AR01056, 26BR01082, 26CR01075, 26DR01167, 26DR01255, 26AR01057, 26BR01083, 26CR01076, 26DR01168, 26ER01022, 26AR01058, 26BR01084, 26CR01079, 26DR01170, 26ER01023, 26AR01059, 26BR01085, 26CR01080, 26DR01171, 26ER01024, 26AR01230, 26BR01087, 26CR01081, 26DR01172, 26ER01029, 26BR01007, 26BR01088, 26CR01082, 26DR01173, 26ER01034, 26BR01035, 26BR01210, 26CR01083, 26DR01174, 26ER01035, 26BR01036, 26CR01014, 26CR01084, 26DR01175, 26ER01037, 26BR01037, 26CR01045, 26CR01085, 26DR01177, 26ER01039, 26BR01039, 26CR01046, 26CR01086, 26DR01180, 26ER01040, 26BR01045, 26CR01047, 26CR01087, 26DR01182, 26ER01042, 26BR01049, 26CR01048, 26CR01088, 26DR01183, 26ER01044, 26BR01051, 26CR01049, 26CR01090, 26DR01184, 26ER01048, 26BR01052, 26CR01051, 26CR01091, 26DR01185, 26ER01049, 26BR01053, 26CR01052, 26CR01092, 26DR01186, 26ER01054, 26BR01057, 26CR01054, 26CR01093, 26DR01187, 26ER01055, 26BR01058, 26CR01056, 26CR01094, 26DR01188, 26ER01056, 26BR01059, 26CR01057, 26CR01206, 26DR01189, 26ER01062, 26BR01060, 26CR01058, 26CR01207, 26DR01190, 26ER01065, 26BR01063, 26CR01059, 26CR01208, 26DR01191, 26ER01066, 26BR01064, 26CR01060, 26CR01209, 26DR01193, 26ER01074, 26BR01065, 26CR01061, 26CR01210, 26DR01194, 26ER01266, 26BR01066, 26CR01064, 26DR01013, 26DR01195, 26ER01268, 26BR01067, 26CR01065, 26DR01158, 26DR01196, 26ER01282, 26BR01069, 26CR01066, 26DR01159, 26DR01197; |
| Distribution | Domestic: Nationwide Distribution; |
| Recall date | Sep 30, 2026 |
| Class | Class II |
| Agency ID | Z-3225-2026 |
| Official notice | U.S. Food and Drug Administration (FDA) enforcement report |
| Data as of, last checked | Nov 21, 2026, 3:48 PM EST |
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- Model Number: 03-2522-1; UDI-DI: 00840861102402 (Bag), 10840861102409 (Case); Lot codes: 26AR01010, 26BR01072, 26CR01067, 26DR01160, 26DR01198, 26AR01011, 26BR01074, 26CR01069, 26DR01161, 26DR01203, 26AR01034, 26BR01076, 26CR01070, 26DR01162, 26DR01206, 26AR01035, 26BR01077, 26CR01072, 26DR01163, 26DR01207, 26AR01046, 26BR01078, 26CR01073, 26DR01165, 26DR01209, 26AR01047, 26BR01079, 26CR01074, 26DR01166, 26DR01210, 26AR01056, 26BR01082, 26CR01075, 26DR01167, 26DR01255, 26AR01057, 26BR01083, 26CR01076, 26DR01168, 26ER01022, 26AR01058, 26BR01084, 26CR01079, 26DR01170, 26ER01023, 26AR01059, 26BR01085, 26CR01080, 26DR01171, 26ER01024, 26AR01230, 26BR01087, 26CR01081, 26DR01172, 26ER01029, 26BR01007, 26BR01088, 26CR01082, 26DR01173, 26ER01034, 26BR01035, 26BR01210, 26CR01083, 26DR01174, 26ER01035, 26BR01036, 26CR01014, 26CR01084, 26DR01175, 26ER01037, 26BR01037, 26CR01045, 26CR01085, 26DR01177, 26ER01039, 26BR01039, 26CR01046, 26CR01086, 26DR01180, 26ER01040, 26BR01045, 26CR01047, 26CR01087, 26DR01182, 26ER01042, 26BR01049, 26CR01048, 26CR01088, 26DR01183, 26ER01044, 26BR01051, 26CR01049, 26CR01090, 26DR01184, 26ER01048, 26BR01052, 26CR01051, 26CR01091, 26DR01185, 26ER01049, 26BR01053, 26CR01052, 26CR01092, 26DR01186, 26ER01054, 26BR01057, 26CR01054, 26CR01093, 26DR01187, 26ER01055, 26BR01058, 26CR01056, 26CR01094, 26DR01188, 26ER01056, 26BR01059, 26CR01057, 26CR01206, 26DR01189, 26ER01062, 26BR01060, 26CR01058, 26CR01207, 26DR01190, 26ER01065, 26BR01063, 26CR01059, 26CR01208, 26DR01191, 26ER01066, 26BR01064, 26CR01060, 26CR01209, 26DR01193, 26ER01074, 26BR01065, 26CR01061, 26CR01210, 26DR01194, 26ER01266, 26BR01066, 26CR01064, 26DR01013, 26DR01195, 26ER01268, 26BR01067, 26CR01065, 26DR01158, 26DR01196, 26ER01282, 26BR01069, 26CR01066, 26DR01159, 26DR01197;
- Quantity
- 1743026 units
- Distribution
- Domestic: Nationwide Distribution;
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC, Waltham, MA
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-07-29
Source: openFDA enforcement record Z-3225-2026 · search recall number Z-3225-2026 in FDA's Enforcement Report.
Copied from the agency record without changes. Data as of Nov 21, 2026, 3:48 PM EST. The agency notice is the final word.
FAQ
What was recalled?
Product: CombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors; Model Number: 03-2522-1; Brand: Fresenius Medical Care Renal Therapies Group, LLC Codes, lots and dates: Model Number: 03-2522-1; UDI-DI: 00840861102402 (Bag), 10840861102409 (Case); Lot codes: 26AR01010, 26BR01072, 26CR01067, 26DR01160, 26DR01198, 26AR01011, 26BR01074, 26CR01069, 26DR01161, 26DR01203, 26AR01034, 26BR01076, 26CR01070, 26DR01162, 26DR01206, 26AR01035, 26BR01077, 26CR01072, 26DR01163, 26DR01207, 26AR01046, 26BR01078, 26CR01073, 26DR01165, 26DR01209, 26AR01047, 26BR01079, 26CR01074, 26DR01166, 26DR01210, 26AR01056, 26BR01082, 26CR01075, 26DR01167, 26DR01255, 26AR01057, 26BR01083, 26CR01076, 26DR01168, 26ER01022, 26AR01058, 26BR01084, 26CR01079, 26DR01170, 26ER01023, 26AR01059, 26BR01085, 26CR01080, 26DR01171, 26ER01024, 26AR01230, 26BR01087, 26CR01081, 26DR01172, 26ER01029, 26BR01007, 26BR01088, 26CR01082, 26DR01173, 26ER01034, 26BR01035, 26BR01210, 26CR01083, 26DR01174, 26ER01035, 26BR01036, 26CR01014, 26CR01084, 26DR01175, 26ER01037, 26BR01037, 26CR01045, 26CR01085, 26DR01177, 26ER01039, 26BR01039, 26CR01046, 26CR01086, 26DR01180, 26ER01040, 26BR01045, 26CR01047, 26CR01087, 26DR01182, 26ER01042, 26BR01049, 26CR01048, 26CR01088, 26DR01183, 26ER01044, 26BR01051, 26CR01049, 26CR01090, 26DR01184, 26ER01048, 26BR01052, 26CR01051, 26CR01091, 26DR01185, 26ER01049, 26BR01053, 26CR01052, 26CR01092, 26DR01186, 26ER01054, 26BR01057, 26CR01054, 26CR01093, 26DR01187, 26ER01055, 26BR01058, 26CR01056, 26CR01094, 26DR01188, 26ER01056, 26BR01059, 26CR01057, 26CR01206, 26DR01189, 26ER01062, 26BR01060, 26CR01058, 26CR01207, 26DR01190, 26ER01065, 26BR01063, 26CR01059, 26CR01208, 26DR01191, 26ER01066, 26BR01064, 26CR01060, 26CR01209, 26DR01193, 26ER01074, 26BR01065, 26CR01061, 26CR01210, 26DR01194, 26ER01266, 26BR01066, 26CR01064, 26DR01013, 26DR01195, 26ER01268, 26BR01067, 26CR01065, 26DR01158, 26DR01196, 26ER01282, 26BR01069, 26CR01066, 26DR01159, 26DR01197;
What is the hazard?
Communication to recommend reviewing the IFU and ensuring that all treatment preparation and handling steps are performed as instructed.
How do I check if mine is affected?
Codes, lots and dates: Model Number: 03-2522-1; UDI-DI: 00840861102402 (Bag), 10840861102409 (Case); Lot codes: 26AR01010, 26BR01072, 26CR01067, 26DR01160, 26DR01198, 26AR01011, 26BR01074, 26CR01069, 26DR01161, 26DR01203, 26AR01034, 26BR01076, 26CR01070, 26DR01162, 26DR01206, 26AR01035, 26BR01077, 26CR01072, 26DR01163, 26DR01207, 26AR01046, 26BR01078, 26CR01073, 26DR01165, 26DR01209, 26AR01047, 26BR01079, 26CR01074, 26DR01166, 26DR01210, 26AR01056, 26BR01082, 26CR01075, 26DR01167, 26DR01255, 26AR01057, 26BR01083, 26CR01076, 26DR01168, 26ER01022, 26AR01058, 26BR01084, 26CR01079, 26DR01170, 26ER01023, 26AR01059, 26BR01085, 26CR01080, 26DR01171, 26ER01024, 26AR01230, 26BR01087, 26CR01081, 26DR01172, 26ER01029, 26BR01007, 26BR01088, 26CR01082, 26DR01173, 26ER01034, 26BR01035, 26BR01210, 26CR01083, 26DR01174, 26ER01035, 26BR01036, 26CR01014, 26CR01084, 26DR01175, 26ER01037, 26BR01037, 26CR01045, 26CR01085, 26DR01177, 26ER01039, 26BR01039, 26CR01046, 26CR01086, 26DR01180, 26ER01040, 26BR01045, 26CR01047, 26CR01087, 26DR01182, 26ER01042, 26BR01049, 26CR01048, 26CR01088, 26DR01183, 26ER01044, 26BR01051, 26CR01049, 26CR01090, 26DR01184, 26ER01048, 26BR01052, 26CR01051, 26CR01091, 26DR01185, 26ER01049, 26BR01053, 26CR01052, 26CR01092, 26DR01186, 26ER01054, 26BR01057, 26CR01054, 26CR01093, 26DR01187, 26ER01055, 26BR01058, 26CR01056, 26CR01094, 26DR01188, 26ER01056, 26BR01059, 26CR01057, 26CR01206, 26DR01189, 26ER01062, 26BR01060, 26CR01058, 26CR01207, 26DR01190, 26ER01065, 26BR01063, 26CR01059, 26CR01208, 26DR01191, 26ER01066, 26BR01064, 26CR01060, 26CR01209, 26DR01193, 26ER01074, 26BR01065, 26CR01061, 26CR01210, 26DR01194, 26ER01266, 26BR01066, 26CR01064, 26DR01013, 26DR01195, 26ER01268, 26BR01067, 26CR01065, 26DR01158, 26DR01196, 26ER01282, 26BR01069, 26CR01066, 26DR01159, 26DR01197;
Where was it distributed?
Domestic: Nationwide Distribution;
Related recalls
- Fresenius Medical Care Renal Therapies Group, LLC: DELFLEX Peritoneal Dialysis Solution with 2 recall
- Fresenius Medical Care Renal Therapies Group, LLC: DELFLEX Peritoneal Dialysis Solution with 1 recall
- Fresenius Medical Care Renal Therapies Group, LLC: DELFLEX Peritoneal Dialysis Solution with 1 recall
- Fresenius Medical Care Renal Therapies Group, LLC: DELFLEX Peritoneal Dialysis Solution with 2 recall
- Fresenius Medical Care Renal Therapies Group, LLC: DELFLEX Peritoneal Dialysis Solution with 4 recall
- Fresenius Medical Care Renal Therapies Group, LLC: 5008X HD/HDF with CLiC Blood Tubing Set Part number 03-5300-3C recall