FDAFDA recall Z-0351-2027 · Nov 11, 2026
Baxter Healthcare Corporation: Operating table column TS7500 MOBIUS recall
Recall found
Answer
Yes, Baxter Healthcare Corporation Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695 is recalled as of Nov 11, 2026. On Nov 11, 2026, FDA announced recall Z-0351-2027 covering 77 units. What to do: follow the official FDA notice. Last checked Nov 17, 2026, 1:52 PM EST.
Product
Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695
Reason for recall
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
What to do
FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.
| Units | 77 units |
|---|---|
| Lot, UPC, model, or component | Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411 |
| Distribution | Worldwide - US Nationwide distribution. |
| Recall date | Nov 11, 2026 |
| Class | Class II |
| Agency ID | Z-0351-2027 |
| Official notice | U.S. Food and Drug Administration (FDA) enforcement report |
| Data as of, last checked | Nov 17, 2026, 1:52 PM EST |
Recall details
- FDA classification
- Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
- Status
- Ongoing
- Codes, lots and dates
- Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411, 108516637, 108544106, 108544116, 108618964, 108463224, 108484405, 108489648, 108511412, 108521436, 108544107, 108604137, 108618965, 108468062, 108484406, 108506316, 108511413, 108521452, 108544110, 108604140, 108468063, 108484408, 108506317, 108516634, 108536776, 108544111, 108604153; UDI-DI number: 00887761974425
- Quantity
- 77 units
- Distribution
- Worldwide - US Nationwide distribution.
- Recalling firm
- Baxter Healthcare Corporation, Deerfield, IL
- Voluntary or mandated
- Voluntary: Firm initiated
- Recall started
- 2026-09-29
Source: openFDA enforcement record Z-0351-2027 · search recall number Z-0351-2027 in FDA's Enforcement Report.
Copied from the agency record without changes. Data as of Nov 17, 2026, 1:52 PM EST. The agency notice is the final word.
Questions
Can I get a refund?
Only if the official notice says so. Follow the remedy on this page and the FDA notice for Z-0351-2027. ConsumerSun does not issue refunds.
How do I know if mine is affected?
Compare Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411 with what you own, along with the product name and the recall date. The official notice is the final list.
Where is the official notice?
This page links to the FDA notice for Z-0351-2027. That notice is the final word. ConsumerSun copies the agency record and does not replace it.
Related recalls
- Baxter Healthcare Corporation: TruSystem 7500 surgical table recall
- Baxter Healthcare Corporation: Mobile column TruSystem 7500 U recall
- Baxter Healthcare Corporation: Operating Table TruSystem 7000 U recall
- Baxter Healthcare Corporation: TruSystem 7000 U Operating Table (dV) recall
- Baxter Healthcare Corporation: Operating Table TruSystem 7000 U (dV) recall
- Baxter Healthcare Corporation: TruSystem 7500 Hybrid Plus (FC) recall