Skip to content

Recall watch for the stuff you own · As an Amazon Associate we earn from qualifying purchases · Some links earn a commissionDisclosure

ConsumerSun

FDAFDA recall Z-0354-2027 · Nov 11, 2026

Baxter Healthcare Corporation: TruSystem 7500 Hybrid Plus (FC) recall

Recall found

Answer

Yes, Baxter Healthcare Corporation TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087 is recalled as of Nov 11, 2026. On Nov 11, 2026, FDA announced recall Z-0354-2027 covering 1 unit. What to do: follow the official FDA notice. Last checked Nov 17, 2026, 1:52 PM EST.

Product

TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087

Reason for recall

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

What to do

FDA enforcement reports don't include consumer instructions. Check the codes below against your packaging. If this is a medicine or device you use, talk to your pharmacist or doctor before you stop or change anything, and contact the company or FDA for what to do with the product.

Recall facts
Units1 unit
Lot, UPC, model, or componentSerial numbers: 107971925; UDI-DI number: 00887761974067
DistributionWorldwide - US Nationwide distribution.
Recall dateNov 11, 2026
ClassClass II
Agency IDZ-0354-2027
Official noticeU.S. Food and Drug Administration (FDA) enforcement report
Data as of, last checkedNov 17, 2026, 1:52 PM EST

Recall details

FDA classification
Class II: use of or exposure to the product may cause temporary or medically reversible health problems, or the chance of serious problems is remote.
Status
Ongoing
Codes, lots and dates
Serial numbers: 107971925; UDI-DI number: 00887761974067
Quantity
1 unit
Distribution
Worldwide - US Nationwide distribution.
Recalling firm
Baxter Healthcare Corporation, Deerfield, IL
Voluntary or mandated
Voluntary: Firm initiated
Recall started
2026-09-29

Source: openFDA enforcement record Z-0354-2027 · search recall number Z-0354-2027 in FDA's Enforcement Report.

Copied from the agency record without changes. Data as of Nov 17, 2026, 1:52 PM EST. The agency notice is the final word.

Questions

Can I get a refund?

Only if the official notice says so. Follow the remedy on this page and the FDA notice for Z-0354-2027. ConsumerSun does not issue refunds.

How do I know if mine is affected?

Compare Serial numbers: 107971925; UDI-DI number: 00887761974067 with what you own, along with the product name and the recall date. The official notice is the final list.

Where is the official notice?

This page links to the FDA notice for Z-0354-2027. That notice is the final word. ConsumerSun copies the agency record and does not replace it.

Related recalls

Watch my items

Get an email when new recalls are posted for the kinds of things you own. Just your first name and email, nothing else.

Unsubscribe any time with one click. We never share your address.